Respironics Everflo Filters are devised to be used with all EverFlo and EverFlo Q Oxygen Concentrators. They extend the lifespan of the unit. Everflo oxygen concentrator filters are FAA approved. Compressor Intake Filter (Inlet Filter):. Processes air as it enters the concentrator.
Comes with clear case for easy internal filter inspection. Should be replaced on an average, every 2 years Micro-Disk Filter:. Internal part and final component of filtering system.
Ensures thoroughly filtered oxygen before delivery. Replaced as necessary, on an average, every 5 years. How To Replace Micro-Disk Filter?
To Remove the Micro-Disk Filter:. Remove the Filter Cover. Remove the Inlet Filter. Remove the Rear Cabinet. Remove the Flow Tubing from the DISS Outlet Fitting and the top connection of the Flow Meter.
Remove the Flow Tubing from the Micro-Disk Filter. Remove the Pressure Tubing from the Pressure sensor located on the Main PCA (OPI Units Only). Remove the other end of the Pressure Tubing from the bottom connection of the Flow Meter and from the Pressure Regulator (OPI Units Only). To Install the Micro-Disk Filter:. Connect the Pressure Tubing to the Pressure Sensor located on the Main PCA (OPI Units Only). Connect the other ends of the Pressure Tubing to the Pressure Regulator and bottom connection on the Flow Meter. Professional cooking study guide answers.
Connect the Flow Tubing. Short Piece should be inserted into the inlet side of the Micro-Disk Filter.
Install the Rear Cabinet. Install the Inlet Filter. Install the Filter Cover.
. EverFlo EverFlo Q 1053621 R05 LZ EN-INTL, DE. Symbol Key. 2 Abbreviations.2 Chapter 1: Introduction.3 Intended Use.3 About Your EverFlo / EverFlo Q.3 Parts of Your Concentrator.3 Accessory Equipment and Replacement Parts.4 Warnings and Cautions.4 Chapter 2: Operating Instructions.
Compliant with the Waste Electrical and Electronic Equipment/Restriction of the Use of European Declaration of Certain Hazardous Substances in Electrical and Conformity Electronic Equipment (WEEE/RoHS) recycling directives Abbreviations Light Emitting Diode Liters per Minute Oxygen Percentage Indicator © 2014 Koninklijke Philips N.V. All rights reserved. Intended Use The EverFlo / EverFlo Q Oxygen Concentrator is intended to provide supplemental oxygen to persons requiring oxygen therapy. The device is not intended to be life supporting or life sustaining.
About Your EverFlo / EverFlo Q The device produces concentrated oxygen from room air for delivery to a patient requiring low flow oxygen therapy. Note: When connected to the UltraFill Oxygen Filling Station, the EverFlo device disables its Low Flow Alarm when a cylinder is in the process of being filled. If you are breathing from the EverFlo at this time and an occlusion occurs in the cannula, the EverFlo will not sense the disruption of the flow of oxygen to you. If you are breathing from the EverFlo device while filling a cylinder, ensure that you place your oxygen cannula in a position to avoid it being kinked or crushed. Application accessories shall include a means to reduce the propagation of fire. The use of accessories, transducers, and cables other than those specified by Philips Respironics may result in increased emissions or decreased immunity of the device. Step 3-B4 Fill your humidifier bottle according to the manufacturer’s instructions.
Mount the filled humidifier on the top of the EverFlo / EverFlo Q device inside the Velcro strap, as shown in the illustration on the right. Tighten the Velcro strap around the bottle and secure it so it is held firmly in place. EverFlo / EverFlo Q User Manual Press the power switch to the On I position. Initially, all the LEDs will illuminate and the audible alert will beep for a few seconds. After that time, only the green LED should remain lit. You can begin breathing from the device immediately even though it typically takes 10 minutes to reach oxygen purity specifications.
Respironics Everflo Instructions
Caution: Excess moisture may impair the proper operation of the device. Cleaning Periodically, use a damp cloth to wipe down the exterior case of the EverFlo / EverFlo Q device. If you use medical disinfectants, be sure to follow the manufacturer’s instructions. EverFlo / EverFlo Q User Manual Chapter 4: Alarms and Troubleshooting Alarm and Indicators The device has an audible alarm and three LED indicators, as shown below. NORMAL Green Power On Indicator Yellow Check System Indicator Red Alarm Indicator Audible Alarm / Colored LED. EverFlo / EverFlo Q User Manual Troubleshooting Guide Problem Why it Happened What to Do Turn the flow rate down to your prescribed level. Wait at least 2 minutes.
Yellow LED is blinking. The Red LED is off and the. EverFlo / EverFlo Q User Manual Chapter 5: Specifications Environmental Operating Transport & Storage Temperature 13 to 32° C -34 to 71° C Relative Humidity 15 to 95%, noncondensing 15 to 95%, noncondensing Altitude 0 to 2286 m Physical Dimensions 58 cm x 38 cm x 24 cm. 1039366, 10365, 10370, 110, 1020014.